Introduction
Ultrasound, as currently practiced in medicine, is a real-time tomgraphic
imaging modality. Not only does it produce real-time tomograms of the
position of reflecting surfaces (internal organs and structures), but
it can be used to produce real-time images of tissue and blood motion.
Theory and Instrumentation
Ultrasound denotes the use of acoustical (sound) waves at frequencies
greater than 20 kHz. Generally, medical ultrasound is performed at frequencies
in the range of 1 MHz. The technique is used to determine the location
of surfaces within tissues by measuring the time interval between the
production of an ultrasonic pulse and the detection of its echo resulting
from the pulse reflected from those surfaces. By measuring the time interval
between the transmitted and detected pulse, we can calculate he distance
between the transmitter and the object. The ultrasound pulses are both
produced and detected by a piezoelectric crystal. The crystal has the
property of changing its physical dimensions in response to an electric
field, and can produce an electric field if its physical shape is changed
mechanically. Thus, ultrasonic compression waves (vibrations) are produced
by applying an oscillating potential across the crystal. The reflected
ultrasound imposes a distortion on the crystal, which in turn produces
an oscillating voltage in the crystal. The same crystal is used for both
transmission and reception.
Doppler Ultrasound
If a structure is stationary, the frequency of the reflected wave will
be identical to that of the impinging wave. A moving structure will cause
a back-scattered signal frequency shifted higher or lower depending on
the structure's velocity toward or away from the sound generator (called
a transducer).
For example, when an impinging sound pulse passes through a blood vessel,
scattering and reflection occurs from the moving red cells. In this process,
small amounts of sound energy are absorbed by each red cell, then re-radiated
in all directions. If the cell is moving with respect to the source, the
back scattered energy returning to the source will be shifted in frequency,
with the magnitude and direction proportional to the velocity of the respective
blood cell. Thus, if we use ultrasound to image the cross-sectional area
of the blood vessel, the volume of blood flow can be calculated from the
area of the vessel and the average velocities of the blood cells.
Clinical Applications
The major use of Doppler ultrasound is the study of the heart and human
carotid artery disease wherein imaging and frequency shift are combined
to produce images of artery and ventricle lumens. The frequency shift
data is used to color the image, showing direction of flow (e.g. carotid
arteries in red and veins in blue). Obstructions to blood flow are readily
evaluated by this method using hand held scanning devices.
In addition to imaging heart valves and blood vessels, ultrasound is the
most convenient and inexpensive method for medical evaluations such as
fetal gender and gallbladder stones. Ultrasound imaging is also being
used for monitoring therapy methods such as hyperthermia, cryosurgery,
drug injections, and as a guide during biopses and catheter placements.